MBU500

508.6W Medical Open-Frame Switching Power Supplies

● Size: 3.00" x 5.00" x 1.38"
● Safety: 60601 3.1, 60601 3.2
● Class I
● Operating Voltage: 90-264 VAC, 47-63 Hz
● Single Output
● Protection: SCP, OLP, OVP, OTP, OCP
● 2MOPP
● High Power Density
● Suitable for Medical/Healthcare Equipment
● 5-Year Warranty

Model No. Output Voltage / FAN Output(VDC) Max. Output Current / FAN Output(A) Max. Output Power(W) Safety Approvals
Model No.:MBU500-105(P) Output Voltage / FAN Output(VDC):12.0 / 12.0 Max. Output Current / FAN Output(A):41.67 / 0.3 Max. Output Power(W):508.6 下载 :More
Model No.:MBU500-106(P) Output Voltage / FAN Output(VDC):15.0 / 12.0 Max. Output Current / FAN Output(A):33.33 / 0.3 Max. Output Power(W):508.6 下载 :More
Model No.:MBU500-107(P) Output Voltage / FAN Output(VDC):19.0 / 12.0 Max. Output Current / FAN Output(A):26.31 / 0.3 Max. Output Power(W):508.6 下载 :More
Model No.:MBU500-108(P) Output Voltage / FAN Output(VDC):24.0 / 12.0 Max. Output Current / FAN Output(A):20.83 / 0.3 Max. Output Power(W):508.6 下载 :More
Model No.:MBU500-109(P) Output Voltage / FAN Output(VDC):30.0 / 12.0 Max. Output Current / FAN Output(A):16.66 / 0.3 Max. Output Power(W):508.6 下载 :More
Model No.:MBU500-110(P) Output Voltage / FAN Output(VDC):36.0 / 12.0 Max. Output Current / FAN Output(A):13.88 / 0.3 Max. Output Power(W):508.6 下载 :More
Model No.:MBU500-111(P) Output Voltage / FAN Output(VDC):48.0 / 12.0 Max. Output Current / FAN Output(A):10.41 / 0.3 Max. Output Power(W):508.6 下载 :More
Type File Name Download
Type:Safety Instructions File Name:Important Safety Instructions Download : Download
Type:Datasheet File Name:SINPRO-MBU500-DATASHEET Download : Download
Type:Instruction Manual File Name:SINPRO-MBU500-Fan kit Instruction Manual(93-41635-01) Download : Download
Type:Instruction Manual File Name:SINPRO-Instruction Manual(MBU500P)_EN Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-cULus(MED3.2) Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-CB(MED3.2) Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-CE(EMC4.1) Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-CB(MED3.1) Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-cULus(MED3.1) Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-UL Download : Download
Type:Safety Certificate File Name:SINPRO-MBU500-TUV(MED3.2) Download : Download

Description

The MBU500 series is a family of open frame medical power supplies that are designed to offer 500W continuous output power with a single output of either 12v, 24v, 36v, or 48v (coming with a universal input of 90-264 Vac) at efficiencies up to 92%. The power modules can operate in an operating temperature range of -40 to +70°C with a 5000M altitude operational ability.
 
The medical power supplies are burn-in test conducted and offer class I operation, low leakage current, overload protection (OLP), overvoltage protection (OVP), over-temperature protection (OTP), and 2 x MOPP (Means of Patient Protection) classification to ensure the safety of the patient, caregiver, and device in which the power supply is installed. The series owns comprehensive protection designs along with all-inclusive safety certifications and conforms to major international safety standards including IEC/ES/EN/CSA C22.2 60601-1. All modules are fully compliant with RoHS directive.
 
The MBU500 switching power supplies are designed for medical applications and are ideal for ultrasound systems, blood chemistry analyzers, medical imaging equipment, patient monitors, etc.
 
To find more applications and product details, please click the contact us button below and experience the customized service.

Electrical Characteristics
● Efficiency: 90~92% (typ.)
● Line Regulation: 1% (max.)
● Load Regulation: 3% (max.)
● Hold-Up Time: 12mS (min.) (with 25 CFM forced air)
● Output Ripple and Noise: 1% (typ.)
 
Environmental

● Operating Temperature: -40 to +70°C
● Storage Temperature: -40 to +85°C
● Operating Humidity: 0~95%
● Storage Humidity: 0~95%
● MTBF: 200,000 calculated hours

*"P" means with standby output : (5V@0.5A) Air convection ; (5V@1A) Forced Air